Catalog Number 10697306
Device Problem Low Test Results (2458)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Manufacturer Narrative
The customer has provided instrument log and sensor files for investigation.The customer has replaced the measurement cartridge and has been reviewing sodium results to confirm proper operation of the instrument.Investigation is ongoing.The cause of this event is unknown.
 
Event Description
The customer reported a discrepant low sodium result on a neonatal patient sample when compared to retesting of the same sample on a laboratory instrument.There is no report of injury due to this event.
 
Manufacturer Narrative
Siemens healthcare diagnostics inc.Has confirmed an issue with rapidpoint 500 systems measurement cartridges (with lactate).This issue has the potential to affect the sodium (na+) sensor, as well as cause a question result ¿-----?¿ error flag for multiple electrolytes on patient samples and quality control.When the na+ sensor is affected, na+ slope calibration errors (d3) will be seen in the recall events log during cartridge initialization, resulting in low na+ values.The maximum sodium bias on internal testing and reported by customers comparing results to other direct method analyzers was <10 mm.To date, the impact to na+ results has been observed in less than 1% of the rapidpoint 500 systems measurement cartridges (with lactate).The question result ¿-----?¿ error flag has been observed on electrolytes.This may occur at any time over the life of the cartridge.Based on the investigation, these issues are due to an electronic noise that is observed on rapidpoint 500 systems measurement cartridges (with lactate).Measurement cartridges without lactate are not affected.Siemens has released a notification "potential for low na+ values and question result error flags" to inform customers of the issue and provide workaround options.Crr# 3002637618-11-01-2023-0004.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RAPIDPOINT 500 BLOOD GAS ANALYZER
Type of Device
RP 500
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown NY
Manufacturer (Section G)
SIEMENS HEALTHCARE DIAGNOSTICS MANUFACTURING LTD.
northern road
chilton industrial estate
sudbury, suffolk CO10 2XQ
UK   CO10 2XQ
Manufacturer Contact
felix akinrinola
2 edgewater drive
norwood, MA 02062
MDR Report Key18144835
MDR Text Key328982995
Report Number3002637618-2023-00097
Device Sequence Number1
Product Code CHL
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K122539
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Physician
Remedial Action Notification
Type of Report Initial,Followup
Report Date 11/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number10697306
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/02/2023
Initial Date FDA Received11/15/2023
Supplement Dates Manufacturer Received11/02/2023
Supplement Dates FDA Received11/30/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Type of Device Usage Reuse
Removal/Correction Number300263761811012023-0004
Patient Sequence Number1
-
-