MAUDE Adverse Event Report: ABBOTT ENDURITY¿ CORE DR, JPD; PULSE GENERATOR, PERMANENT, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT ENDURITY¿ CORE DR, JPD; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number PM2152
Device Problem Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2023
Event Type  malfunction  
Event Description
It was reported that the data on the device appeared to be corrupted on the merlin programmer prior to the implant procedure.The device was not used and a new device was implanted to resolve the event.The patient was in stable condition.
 
Manufacturer Narrative
The reported event of data problem was confirmed.The device voltage was above elective replacement indicator (eri) voltage level upon receipt.Analysis of the device image indicated device was within normal range of operation.Further analysis indicated strange inscriptions noted in the field was due to patient information data having non-zero value on a shipped set device.This resulted into the data problem reported by the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURITY¿ CORE DR, JPD
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18145090
MDR Text Key328223141
Report Number2017865-2023-52576
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
Reporter Country CodeIT
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM2152
Device Lot NumberP000173092
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/15/2023
Supplement Dates Manufacturer Received02/08/2024
Supplement Dates FDA Received02/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-