BIOSENSE WEBSTER INC WEBSTER CS CATHETER WITH AUTO ID TECHNOLOGY; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Catalog Number D135303 |
Device Problem
Contamination /Decontamination Problem (2895)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/30/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by biosense webster inc., or its employees that the report constitutes an admission that the product, biosense webster inc., or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.#(b)(4).
|
|
Event Description
|
It was reported that a patient underwent a cardiac ablation procedure with a webster cs catheter with auto id technology and a odd silicone/plastic attachment at the tip of the catheter was found.It was reported they were struggling to advance webster cs catheter with auto id technology and upon pulling it out, they noticed an odd silicone/plastic attachment at the tip of the catheter.There was no patient consequence.Multiple attempts have been made to obtain clarification to this complaint.However, no further information has been made available.
|
|
Manufacturer Narrative
|
On 30-jan-2024, the bwi product analysis lab received the complaint device for evaluation.The product analysis has begun but is not completed at this time.When the investigational analysis has been completed, a supplemental 3500a report will be submitted.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# (b)(4).
|
|
Manufacturer Narrative
|
It was reported that a patient underwent a cardiac ablation procedure with a webster cs catheter with auto id technology and a odd silicone/plastic attachment at the tip of the catheter was found.It was reported they were struggling to advance webster cs catheter with auto id technology and upon pulling it out, they noticed an odd silicone/plastic attachment at the tip of the catheter.There was no patient consequence.Device evaluation details: the device was returned to biosense webster inc (bwi) for evaluation and the evaluation has been completed.A visual inspection evaluation of the returned device was performed following bwi procedures.Visual analysis of the returned sample revealed a plastic foreign material attached to the tip of the device.For this reason, a fourier-transform infrared spectroscopy (ft-ir) analysis was requested and it was concluded that the foreign material is mainly composed of biological-based material.Presumably a tissue or fluid from the patient.A manufacturing record evaluation was performed for the finished device batch number, and no internal actions were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's ref.# pc-(b)(4).
|
|