This device is classified as import for export, therefore 510k is not applicable.Model eb-1575k is available in the usa with a 510k number k131028.The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb shadow in image.Based on the result, we concluded that it was caused due to the moisture condensation in the ccd module with drive pcb.In addition, our technician confirmed that the lcb (light carrying bundle) broken, the insertion flexible tube buckled, the light guide cable buckled, the remote control buttons cut, and the distal body worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|