Catalog Number 2112550
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2023
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As reported by the user facility: brief inquiry description: it was reported that after initial setup the apex transfer set had a black particulate in port 2 macro of the tubing just before the manifold.No patient involvement.
 
Manufacturer Narrative
This report has been identified as b.Braun medical internal report number (b)(4).One (1) used sample without packaging was returned to the manufacturer for evaluation.Through visual examination, an embedded particulate on the outside of the tube that inserts into the manifold was observed.The sample is not within specification; the reported defect is confirmed.A review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.Incidents of this nature are attributed to operator oversight during the assembly of the product.As a result of this occurrence, a formal awareness training session was conducted with all applicable personnel involved in the packaging and inspection of this product.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
B.BRAUN
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
B. BRAUN MEDICAL INC.
901 marcon blvd.
allentown PA 18109
Manufacturer Contact
jonathan severino
901 marcon blvd.
allentown, PA 18109
4847197287
MDR Report Key18147501
MDR Text Key328248784
Report Number2523676-2023-00737
Device Sequence Number1
Product Code LHI
UDI-Device Identifier04046964656101
UDI-Public(01)04046964656101
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K151423
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2112550
Device Lot Number0061901604
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/30/2023
Initial Date FDA Received11/16/2023
Supplement Dates Manufacturer Received10/30/2023
Supplement Dates FDA Received03/28/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/27/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-