• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER Back to Search Results
Catalog Number 175812
Device Problem Inaccurate Flow Rate (1249)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
It was reported that the district nurse visited for catheter change due to urinary tract infection, observations and to obtain a sample.Antibiotics commenced that morning.The patient had history of multiple sclerosis (ms), an enlarged prostate and a large bladder stone, recurrent urinary tract infections.The previous catheter was removed with no issues, some sediment was noted in urine in leg bag.A new catheter inserted, had some resistance and was unable to advance the catheter any further.The balloon inflated and then waited for draining.While waiting encouraged fluids and got patient to stand up.After around 30 minutes no urine had drained.The patient was not in discomfort.Stated that the patient did not had usual pain response, so it was not a good indicator of anything.They tried to deflate the balloon and reposition catheter.However when attaching syringe, it did not spontaneously fill as usual, so they pulled back and only 2ml was in syringe.They tried a different 10ml syringe however faced the same issue.They re-inserted the 2mls and phoned back to office for advice.The clinic nurse checked catheter policy and advice was to leave a 10ml syringe on for a few minutes to see if it would spontaneously fill, this was unsuccessful.Advised to change leg bag, also unsuccessful.
 
Manufacturer Narrative
The reported event was inconclusive because no sample was returned.A potential root cause for this failure could be ¿lumen wall thickness undersized ¿.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use.If package is opened or if any imperfection or surface deterioration is observed, do not use.Ensure catheter and drainage bag tubing is kink free and drainage bag is positioned below the bladder to ensure urine is flowing freely.Periodic inspection of this system should be made to ensure that urine is flowing freely.If a standing column of urine is observed, check for correct positioning of bag and then for a physical obstruction.If correct positioning or removal of physical obstruction does not allow free flow, further investigation should be taken in line with local protocol." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® LUBRI-SIL® ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer (Section G)
C.R. BARD INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
Manufacturer Contact
xeeroy rada
8195 industrial blvd
covington 30014
7707846100
MDR Report Key18147735
MDR Text Key328262694
Report Number1018233-2023-08157
Device Sequence Number1
Product Code EZL
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K984084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 10/31/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number175812
Device Lot NumberNGGW3018
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/23/2023
Initial Date FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-