• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH HEATER/COOLER 3T 110V/60 HZ; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-85
Device Problem Insufficient Cooling (1130)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/21/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t was not cooling down on the patien side more then 25 degrees post procedure.There is no report of any patient injury.
 
Manufacturer Narrative
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in usa.A livanova field service engineer reported also the customer complained of getting a press switch tripped (e27) and display button where not responding and needed to be replaced.The display keyboard, the processor board and patient circuit bridge were replaced.All components have been inspected and tested according unit returned to customer.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: complaints database analysis revealed that no similar event on this device occurred since its installation in 2020.Based on the collected information, the root cause of the reported event of insufficient cooling can be traced back to a defective patient bridge.The wearing of electro-mechanical device can be originated by the specific use condition at customer site and may have contributed to the event, however there is no concerning trend for this kind of failure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER/COOLER 3T 110V/60 HZ
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18148180
MDR Text Key328269863
Report Number9611109-2023-00609
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817900528
UDI-Public010403381790052811201008
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial,Followup
Report Date 02/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-85
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/21/2023
Initial Date FDA Received11/16/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/08/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-