The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The device evaluation found the injector could discharge liquid normally, push and pull the handle , the device could be normally telescopic, there was no cease in the sheath of the injector, and there was crystals in distal end of the sheath.Based on the results of the investigation, it is likely that the following led to the malfunction: the cause of the occurrence could not be determined because the event could not be confirmed, the subject device had no irregularity and the review result of the dhr was not problematic.A device history review revealed no anomaly with the event-related items.The record includes the following: process inspection sheet, quality inspection sheet and nonconforming product report.Although the root cause of this event could not be determined, we will continue to monitor trends and take appropriate actions as necessary.This issue is addressed in the instructions for use (ifu): before use, prepare and inspect the instrument as instructed below.Should the any irregularity be observed, do not use the instrument; use a spare instead.Damage or irregularity may compromise patient or user safety, for example: posing an infection-control risk, causing tissue irritation, perforation, bleeding or mucous membrane damage, and may result in more severe equipment damage.Straighten out the instrument before inspecting it.The instrument can be damaged if it is coiled while the handle is operated.Do not coil the insertion portion with a diameter of less than 15 centimeters.This could damage the insertion portion.Before use, confirm that the needle and the insertion portion are not damaged.If any abnormalities such as significant deformations or excessive bends are found, do not use the instrument.Otherwise, it may cause perforation, bleeding, mucous membrane damage.Confirm that the needle extends in the endoscopic image are normal.If any abnormalities are found, do not use the instrument.Otherwise, it may cause perforation, bleeding or mucous membrane damage.When inserting the instrument into the endoscope, retract the needle into the tube, hold the instrument close to the biopsy valve, and keep it as straight as possible relative to the biopsy valve.Otherwise, the instrument could be damaged.Stop using the instrument if the insertion portion bends excessively during use.This could result in malfunction, such as failing to extend the needle or inject a fluid.Do not push the slider abruptly, otherwise the needle will be rapidly extended from the distal end of the tube.This could result in patient injury, such as perforation, bleeding or mucous membrane damage.It could also damage the instrument.Olympus will continue to monitor the field performance of this device.
|