Catalog Number 130742203
Device Problem Naturally Worn (2988)
Patient Problems Pain (1994); Unspecified Tissue Injury (4559)
Event Date 11/01/2023
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Patient was revised due to painful shoulder with external rotation.Scapula notching suspected.Notching was occurred at inferior edge.Poly wear was also indicated.Surgeon replaced glenosphere without 42+6 ecc ball and put in a new 42+3 poly liner.Surgical delay unknown.Doi: (b)(6) 2022.Dor: nov 1, 2023.Affected side: left shoulder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DXTEND STAND PE CUP D42 +3MM
Type of Device
DELTA XTEND IMPLANTS : SHOULDER HUMERAL CUP
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18148938
MDR Text Key328286566
Report Number1818910-2023-23433
Device Sequence Number1
Product Code HSD
UDI-Device Identifier10603295027539
UDI-Public10603295027539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K192855
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/16/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number130742203
Device Lot Number5417455
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2023
Initial Date FDA Received11/16/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
XTND GLENO D42MM +2MM
Patient Outcome(s) Required Intervention;
Patient Age74 YR
Patient SexMale
Patient Weight69 KG
-
-