ST. JUDE MEDICAL ADVISOR¿ HD GRID MAPPING CATHETER, SENSOR ENABLED¿; CATHETER, ELECTRODE RECORDING, OR PROBE, ELECTRODE RECORDING
|
Back to Search Results |
|
Model Number D-AVHD-DF16 |
Device Problems
Fracture (1260); Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/10/2023 |
Event Type
malfunction
|
Event Description
|
Related manufacture ref: 3005334138-2023-00540.During an atypical atrial flutter ablation procedure, exposed wires were noted on the catheter.After completing the map in right atrium (ra), it was understood that the arrhythmia origin was in the left atrium (la).Transseptal puncture was then performed with swartz lamp introducer and brk-1 xs needle.After the puncture, the hd grid was placed in a long sheath to bring it in the la, however, difficulties were noted removing it from the short introducer (boston scientific, 8f).It was finally able to be removed but was damaged with exposed wires.
|
|
Manufacturer Narrative
|
Additional information: g3, h2, h3 one image was submitted for evaluation to product performance engineering; no device components were returned.The photo appeared to show an advisor hd grid paddle with damage.The paddle assembly appeared to be damaged, exposing the nitinol frame.Visual inspection was based solely upon a review of the photographs provided.The device was not returned.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Arten600124477 ver.B advisor hd grid diagnostic catheter instructions for use (ifu) states, ¿insert the distal tip section of catheter into an 8.5 f minimum introducer (not included) using the insertion tool¿.The cause of the paddle damage and removal difficulty is consistent with not following the instructions for use.
|
|
Search Alerts/Recalls
|
|
|