Model Number 103
Device Problem Inappropriate or Unexpected Reset (2959)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/06/2023
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any defects ¿or malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported, that upon interrogating the patient device, error code 1 (hardware reset) was seen indicating the device was disabled and not supplying stimulation.Impedance, and output were not showing up on the diagnostic test despite the battery life being ok.Device history records were reviewed for the patient implanted generator and the device passed all functional specifications.Ap chest x-ray was received and reviewed.The x-rays were provided after report of generator reset.The generator placement appeared to be per labeling.Based on the images provided, the feedthrough wires were intact, and the connector pin appear to be fully inserted past the backend of the connector block.The visible portion of the lead was reviewed, but due to the scope of the image the strain relief and tie downs could not be assessed.A portion of the lead was visualized behind the generator.Based on the x-rays received, the cause of the generator reset is still unknown.However, any fractures or microfractures on the portions of the lead that were not visible cannot be ruled out.No known relevant surgical intervention has occurred to date.No other relevant information has been received to date.
 
Event Description
It was reported that the patient underwent a replacement surgery.The explanted generator has not been returned to date.No other relevant information has been received to date.
 
Event Description
Additional information was received confirming the date of explant.No other relevant information was received to date.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 103
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key18160509
MDR Text Key328411634
Report Number1644487-2023-01652
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750030
UDI-Public05425025750030
Combination Product (y/n)N
Reporter Country CodeCO
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 05/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date04/25/2019
Device Model Number103
Device Lot Number204126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Event Location Other
Initial Date Manufacturer Received 10/30/2023
Initial Date FDA Received11/17/2023
Supplement Dates Manufacturer Received03/15/2024
04/24/2024
Supplement Dates FDA Received04/09/2024
05/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/24/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age8 YR
Patient SexFemale
-
-