Model Number NVM5MULTI
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/26/2023
Event Type  malfunction  
Event Description
It was reported that during an anterior cervical discectomy and fusion case at c3-c4 using neuromonitoring, motor evoke potential (mep) and somatosensory evoke potential (ssep) measurement waveforms were inconsistent.There was no patient impact was as a result of the event.
 
Manufacturer Narrative
The device was returned to the manufacturer for evaluation.Although no issue was found in relation to the reported ssep measurement waveforms inconsistency, service tested the system and was able to identify an issue with the mep stimulation pulse parameters.The stimulation circuit board was replaced to address the issue.While the entirety of the reported event was unable to be confirmed, based on the information obtained, the root cause of the reported inconsistent mep monitoring issue can be attributed to an internal circuit board failure.Labeling review: "¿warnings, cautions and precautions.¿if system data acquisition seems inaccurate or if the software application does not initiate or malfunctions during use, and recommended steps to restore the system are not successful, abort use of the system." ".While the nvm5 system is designed to assist in the electromyographic location of spinal nerves in proximity to the surgical site, it is not intended to take the place of thorough knowledge of spinal anatomy and appropriate surgical technique, nor should the information provided by the system be construed as definitive indicators of nerve location.Such factors as the distance from the nerve, the position and placement of electrodes, individual muscle and/or nerve responses, the proximity and strength of sources of electrical interference, and other patient and environmental factors, may influence the operation.If, in the judgment of the clinician, this resistance is sufficient to preclude proper placement of instruments, the procedure should be suspended." ".Pre-operative warnings the methods of use of instruments are to be determined by the user¿s experience and training in surgical procedures.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NUVASIVE NEXT GENERATION NVM5 SYSTEM
Type of Device
NEUROSURGICAL NERVE LOCATOR
Manufacturer (Section D)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer (Section G)
NUVASIVE, INC.
7475 lusk boulevard
san diego CA 92121
Manufacturer Contact
geoff gannon
7475 lusk boulevard
san diego, CA 92121
MDR Report Key18160963
MDR Text Key328415915
Report Number2031966-2023-00263
Device Sequence Number1
Product Code PDQ
UDI-Device Identifier00887517089175
UDI-Public887517089175
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162313
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 11/17/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberNVM5MULTI
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/26/2023
Initial Date FDA Received11/17/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/21/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-