The device was returned to the manufacturer for evaluation.Although no issue was found in relation to the reported ssep measurement waveforms inconsistency, service tested the system and was able to identify an issue with the mep stimulation pulse parameters.The stimulation circuit board was replaced to address the issue.While the entirety of the reported event was unable to be confirmed, based on the information obtained, the root cause of the reported inconsistent mep monitoring issue can be attributed to an internal circuit board failure.Labeling review: "¿warnings, cautions and precautions.¿if system data acquisition seems inaccurate or if the software application does not initiate or malfunctions during use, and recommended steps to restore the system are not successful, abort use of the system." ".While the nvm5 system is designed to assist in the electromyographic location of spinal nerves in proximity to the surgical site, it is not intended to take the place of thorough knowledge of spinal anatomy and appropriate surgical technique, nor should the information provided by the system be construed as definitive indicators of nerve location.Such factors as the distance from the nerve, the position and placement of electrodes, individual muscle and/or nerve responses, the proximity and strength of sources of electrical interference, and other patient and environmental factors, may influence the operation.If, in the judgment of the clinician, this resistance is sufficient to preclude proper placement of instruments, the procedure should be suspended." ".Pre-operative warnings the methods of use of instruments are to be determined by the user¿s experience and training in surgical procedures.".
|