• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITALS MONITOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH EARLYVUE VS30 VITALS MONITOR Back to Search Results
Model Number 863380
Device Problem No Audible Prompt/Feedback (2282)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Event Description
It was reported that the device had a speaker malfunction.It is unknown if the device was in use at time of event, and there was no adverse event reported.
 
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.
 
Event Description
It was reported that the device had a speaker malfunction.The device was not in use on a patient at the time of event, there was no patient involvement.
 
Manufacturer Narrative
The device was returned through the defoa process to the factory for further evaluation.The engineer performed an evaluation/analysis and reproduced the failure.The engineer found that one loudspeaker cable was no longer plugged in, and the second loudspeaker was okay.It was determined to be a mounting failure.The device was scrapped, and good parts were reused.The customer was provided a replacement device to resolve the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EARLYVUE VS30 VITALS MONITOR
Type of Device
EARLYVUE VS30 VITALS MONITOR
Manufacturer (Section D)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM  71034
Manufacturer (Section G)
PHILIPS MEDIZIN SYSTEME BÖBLINGEN GMBH
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
Manufacturer Contact
hauke schik
hewlett-packard-str. 2,
b1-3/d6
boblingen 71034
GM   71034
7031463203
MDR Report Key18161356
MDR Text Key328419342
Report Number9610816-2023-00595
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838091412
UDI-Public00884838091412
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190624
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number863380
Device Catalogue Number863380
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/19/2023
Initial Date FDA Received11/17/2023
Supplement Dates Manufacturer Received12/14/2023
Supplement Dates FDA Received01/08/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/04/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-