Model Number 863380 |
Device Problem
No Audible Prompt/Feedback (2282)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/19/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the device had a speaker malfunction.It is unknown if the device was in use at time of event, and there was no adverse event reported.
|
|
Manufacturer Narrative
|
A follow up report will be submitted once the investigation is complete.
|
|
Event Description
|
It was reported that the device had a speaker malfunction.The device was not in use on a patient at the time of event, there was no patient involvement.
|
|
Manufacturer Narrative
|
The device was returned through the defoa process to the factory for further evaluation.The engineer performed an evaluation/analysis and reproduced the failure.The engineer found that one loudspeaker cable was no longer plugged in, and the second loudspeaker was okay.It was determined to be a mounting failure.The device was scrapped, and good parts were reused.The customer was provided a replacement device to resolve the issue.
|
|