On october 23, 2023, invacare was made aware of an event that occurred in finland involving a dolomite futura 520 rollator.Invacare is filing this medwatch in an abundance of caution due to the dolomite futura 520 rollator was also sold in the u.S by clarke healthcare.This device was manufactured at invacare rea ab in sweden in 2022.The dealer evaluated the rollator and noted the wheels are worn and the left brake slides, otherwise, the device appeared fine.Neither of the findings appeared to cause or contribute to the incident.At this time, it appears that the incident was caused by the wheel getting stuck in a groove in the road causing the user to fall over the rollator.The device has not been returned to invacare.The dolomite futura/dolomite symphony rollator user manual (1543649-d) states the walking aid is suitable for outdoor use on even ground.The user manual also recommends to ensure performance and safety, to inspect the product regularly which includes the wheels¿ tightness, wear and damage.
|