• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. UNKNOWN CUP; PROTHESIS, HIPS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. UNKNOWN CUP; PROTHESIS, HIPS Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Unintended Movement (3026)
Patient Problem Metal Related Pathology (4530)
Event Date 11/02/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).G2: foreign: australia.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported a patient with hip dysplasia had a right total hip arthroplasty.Subsequently, the patient underwent a revision surgery ten years post implantation, due to the neck of the arcos proximal body was impinging on the cup, causing metallosis.No additional information has been provided.
 
Event Description
No additional event information to report at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.D10: unknown screw unknown.No product was returned, or pictures provided; visual and dimensional evaluations could not be performed.Part and lot identification are necessary for review of device history records, neither were provided.Complaint history review cannot be performed without product identification.Medical records were provided and reviewed by a health care professional.Review of the available records identified the following: an initial right tha was performed.Subsequently, a revision occurred due to the neck of the arcos body impinging on the cup causing metallosis.Records were not provided for the revision.Radiographs were provided and reviewed by a health care professional.Review of the available records identified the following: 1.Right tha with long stem femoral component.Oversized acetabular cup with short femoral neck causing impingement on frog leg views.2.Screw fragment proximal femoral diaphysis.A definitive root cause cannot be determined.This complaint was confirmed based on the provided x-rays.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN CUP
Type of Device
PROTHESIS, HIPS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18162637
MDR Text Key328430469
Report Number0001822565-2023-03275
Device Sequence Number1
Product Code JDI
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/14/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberUNKNOWN CUP
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/02/2023
Initial Date FDA Received11/17/2023
Supplement Dates Manufacturer Received02/13/2024
Supplement Dates FDA Received02/14/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Treatment
SEE H10 NARRATIVE.
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age56 YR
Patient SexFemale
-
-