Section d product information has been updated, as well as g1 manufacturing site and g4.Calibration was last performed on 30-aug-2023.The qc recovery data provided was acceptable.The alarm trace showed clot detection, cuvette buffer slide empty, and not enough fluid alarms.The field service engineer (fse) found that the waste pump was defective.The fse replaced the waste pump, replaced a cracked splash guard on the sample probe wash station, primed the system, and performed precision testing.The customer performed calibration and qc successfully.The service maintenance actions performed by the fse resolved the issue.No further issues were reported after the service visit.
|