• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problems Display or Visual Feedback Problem (1184); Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/03/2023
Event Type  malfunction  
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation (procep) became aware that during the aquablation procedural set-up, while priming the aquabeam handpiece, the aquabeam robotic system generated an "e02 - console error" message.Multiple troubleshooting attempts were made with no success.As a result, the aquablation procedure was aborted.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
Component code = 4756 - aquabeam console, a reusable component of the aquabeam robotic system, controls the functionality of the high-velocity waterjet delivered by the aquabeam handpiece.Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Manufacturer Narrative
The aquabeam console was returned for investigation.No visual defects or anomalies were observed.Functional testing confirmed the reported event.Upon further testing, a layer of oil was observed on the encoder lens and sensor, which could potentially cause the reported e02/e03 errors.The root cause of the reported event was determined to be design-related.The issue has been addressed as part of procept's quality management system.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam console / lot number 22c04474 was conducted,which confirmed that there were no non-conformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported failure.The review indicated that the device met all design and manufacturing specifications when released for distribution.The current user manual um0101-00 rev.F, aquabeam robotic system user manual, us, english was reviewed.Table 5 system detected errors and faults e02 - console error release foot pedal and click x.If error persists, turn off and turn on console.E03 - console error release foot pedal and click x.If error persists, turn off and turn on console.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech drive
san jose, CA 95134
6502327291
MDR Report Key18164690
MDR Text Key328969068
Report Number3012977056-2023-00222
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/16D20221122C
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2023
Initial Date FDA Received11/17/2023
Supplement Dates Manufacturer Received01/17/2024
Supplement Dates FDA Received02/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/22/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-