• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: RESPIRONICS, INC. DREAMSTATION AUTO BIPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. DREAMSTATION AUTO BIPAP; VENTILATOR, NON-CONTINUOUS (RESPIRATOR) Back to Search Results
Model Number DSX700S11F
Device Problems Failure to Deliver (2338); Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/19/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device not returned to manufacturer.
 
Event Description
The manufacturer received information alleging a malfunction with dreamstation auto bipap.The patient reports the device is not blowing any air and this has been going on for a month.Troubleshooting was attempted.The patient further reports the device stops running in the middle of the night.The device will turn back on but they don't know why it stops blowing air.There are no error messages seen and troubleshooting was done.The patient reports the device is displaying the correct bipap temperature 15/8 and there are not rips or tears in the seal.The patient states that they have not been using the prescribed bipap pressure.They used it set on cpap mode because the device does not work in bipap mode.The patient further reports that they visited the dme last thursday and they already checked the device.Everything was set up to the correct settings, changed the auto off feature to off, but still having same issue and was advised that the machine is defective.There is no allegation of serious or permanent harm or injury.No medical intervention was required by the patient.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO BIPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18166040
MDR Text Key328512780
Report Number2518422-2023-31010
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberDSX700S11F
Device Catalogue NumberDSX700S11F
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/19/2023
Initial Date FDA Received11/17/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/11/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-