A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during the aquablation procedural set-up, the aquabeam robotic system generated an "e22 - motorpack error" message.A second aquabeam handpiece was used, but the error persisted.A third aquabeam handpiece was used, but the error persisted.As a result, the aquablation procedure was aborted and converted to transurethral ultrasound-guided laser-induced prostatectomy (tulip) therapy.There were no adverse health consequences to the patient due to this event.
|
The aquabeam handpiece was returned for an investigation into the reported event.No visual defects or anomalies were observed with the handpiece.Functional testing was performed on the reported handpiece, but no issues were observed.However, during priming, the handpiece pump cartridge began to leak.Upon further testing, the handpiece signals were observed to be within the nominal range.The reported event was confirmed through log files.The delrin seal was observed under magnification, revealing a large crack around the seal.However, the reason for the crack is still unknown.The root cause for the reported failure was undeterminable; however, the root cause for the leaking cartridge is attributed to the supplier.The aquabeam robotic system's treatment logs file was reviewed, which confirmed the reported event.The system triggered multiple e22 errors during the procedural setup.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 23c00935 was conducted, which confirmed that there was one (1) non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were reworked to address the nonconformance.Upon re-inspection, the lot met all required specifications and was then deemed acceptable to be released for distribution per device specifications.The aquabeam robotic system user manual, (b)(4) rev.G, states the following: table 5: system detected errors and faults.E22 - motorpack error.Release the foot pedal and click x.1) if the error persists, reconnect the handpiece to the motorpack.2) if the error continues, replace handpiece.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|