• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PROCEPT BIOROBOTICS CORPORATION AQUABEAM ROBOTIC SYSTEM; FLUID JET REMOVAL SYSTEM Back to Search Results
Catalog Number AB2000
Device Problem Visual Prompts will not Clear (2281)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/30/2023
Event Type  malfunction  
Manufacturer Narrative
Root cause of reported event has not yet been established.Investigation by manufacturer is currently in-process.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Event Description
A male patient underwent an aquablation procedure for symptomatic benign prostatic hyperplasia (bph).Procept biorobotics corporation became aware that during the aquablation procedural set-up, the aquabeam robotic system generated an "e22 - motorpack error" message.A second aquabeam handpiece was used, but the error persisted.A third aquabeam handpiece was used, but the error persisted.As a result, the aquablation procedure was aborted and converted to transurethral ultrasound-guided laser-induced prostatectomy (tulip) therapy.There were no adverse health consequences to the patient due to this event.
 
Manufacturer Narrative
The aquabeam handpiece was returned for an investigation into the reported event.No visual defects or anomalies were observed with the handpiece.Functional testing was performed on the reported handpiece, but no issues were observed.However, during priming, the handpiece pump cartridge began to leak.Upon further testing, the handpiece signals were observed to be within the nominal range.The reported event was confirmed through log files.The delrin seal was observed under magnification, revealing a large crack around the seal.However, the reason for the crack is still unknown.The root cause for the reported failure was undeterminable; however, the root cause for the leaking cartridge is attributed to the supplier.The aquabeam robotic system's treatment logs file was reviewed, which confirmed the reported event.The system triggered multiple e22 errors during the procedural setup.A review of the device history record (dhr) ab2000-b / serial number (b)(6) and aquabeam handpiece / lot number 23c00935 was conducted, which confirmed that there was one (1) non-conformance issued to this lot during the manufacturing process that could potentially be related to the reported event.The affected units within the lot were reworked to address the nonconformance.Upon re-inspection, the lot met all required specifications and was then deemed acceptable to be released for distribution per device specifications.The aquabeam robotic system user manual, (b)(4) rev.G, states the following: table 5: system detected errors and faults.E22 - motorpack error.Release the foot pedal and click x.1) if the error persists, reconnect the handpiece to the motorpack.2) if the error continues, replace handpiece.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AQUABEAM ROBOTIC SYSTEM
Type of Device
FLUID JET REMOVAL SYSTEM
Manufacturer (Section D)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer (Section G)
PROCEPT BIOROBOTICS CORPORATION
150 baytech drive
san jose CA 95134
Manufacturer Contact
doria esquivel
150 baytech drive
san jose, CA 95134
MDR Report Key18166806
MDR Text Key328489529
Report Number3012977056-2023-00221
Device Sequence Number1
Product Code PZP
UDI-Device IdentifierB614AB20001
UDI-Public+B614AB20001/$$719C00725/16D20190719Y
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
DEN170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberAB2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/03/2023
Initial Date FDA Received11/18/2023
Supplement Dates Manufacturer Received01/12/2024
Supplement Dates FDA Received02/02/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient SexMale
-
-