INSIGHTEC LTD. EXABLATE 4000; MR GUIDED FOCUSED ULTRASOUND SYSTEM Back to Search Results
Model Number 4000
Device Problems Adverse Event Without Identified Device or Use Problem (2993); Insufficient Information (3190)
Patient Problem Burn(s) (1757)
Event Date 10/20/2023
Event Type  Injury  
Manufacturer Narrative
The cause of this reported incident is still under investigation.
 
Event Description
Patient experienced minor scalp burn and blistering following focused ultrasound treatment.
 
Manufacturer Narrative
All treatment parameters were within specifications, treatment flow was as expected, no malfunction in the device was found and no new risks were recognized.The physician assessed that this event will be transient.
 
Event Description
The patient underwent essential tremor treatment.During treatment the patient complained of a burning sensation on the right side of the head/scalp.The minor blistering was visible outside of the treatment area.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXABLATE 4000
Type of Device
MR GUIDED FOCUSED ULTRASOUND SYSTEM
Manufacturer (Section D)
INSIGHTEC LTD.
5 nahum heth street
tirat carmel, 39120
IS  39120
Manufacturer (Section G)
INSIGHTEC LTD.
5 nachum heth
tirat carmel, 39120
IS   39120
Manufacturer Contact
may zimmerman
5 nachum heth
tirat carmel, 39120
IS   39120
MDR Report Key18167069
MDR Text Key328481069
Report Number9615058-2023-00031
Device Sequence Number1
Product Code POH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150038
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/08/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Model Number4000
Device Catalogue NumberSYS940002
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/20/2023
Initial Date FDA Received11/19/2023
Supplement Dates Manufacturer Received10/20/2023
Supplement Dates FDA Received02/08/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Disability;
Patient SexFemale
-
-