Model Number 92346
Device Problem Loss of Osseointegration (2408)
Patient Problems Unspecified Infection (1930); Inadequate Osseointegration (2646)
Event Type  Injury  
Event Description
Per the clinic, the patient experienced infection at the implant site and a loss of osseointegration resulting in fixture loss.The patient was treated with oral antibiotics (specific date and duration not reported).Additional information has been requested but has not been made available as of the date of this report.
 
Manufacturer Narrative
This report is submitted on november 20, 2023.
 
Manufacturer Narrative
Per the clinic, the patient underwent skin revision surgery to excise the excess skin (specific date not reported).The reported adverse event is associated with a returned device; however, the provided clinical information was reviewed by the manufacturer and no specific device analysis is deemed necessary at this time.Previous product examinations have not showed any relationship between a product geometrical deviation and the reported clinical complication.Additionally, there are no indications that a product failure has contributed to the reported issue.Adverse skin reactions around the baha abutment, ranging from slight redness to infected tissue, are common complications with baha implants, and can occur at any time following implantation.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.Implant loss with baha implants can occur at any time following implantation.Known reasons for implant loss include lack of adequate bone quality/quantity, trauma, infection, generalised diseases and surgical complication.The report frequency for these complications is being monitored under cbas complaint and medical device reporting data monitoring plan and the status is updated on a monthly basis.This event is added to this monitoring.This report is submitted on december 26, 2023.
 
Manufacturer Narrative
Per the clinic, the patient underwent silver nitrate cauterization in order to remove excess skin (specific date not reported).This report is submitted on march 19, 2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIA300 IMPLANT 4MM W ABUTMENT 9MM
Type of Device
COCHLEAR BAHA CONNECT SYSTEM
Manufacturer (Section D)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW  435 33
Manufacturer (Section G)
COCHLEAR BONE ANCHORED SOLUTIONS AB
konstruktionsvägen 14
mölnlycke, 435 3 3
SW   435 33
Manufacturer Contact
nabila zahari
unit ug-1, vertical podium
no. 8 jalan kerinchi,
kuala lumpur, wilayah persekutuan 59200
MY   59200
MDR Report Key18167902
MDR Text Key328475970
Report Number6000034-2023-03933
Device Sequence Number1
Product Code LXB
UDI-Device Identifier09321502019767
UDI-Public(01)09321502019767(10)1563387(17)270718
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K100360
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/21/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model Number92346
Device Catalogue Number92346
Device Lot NumberCOH1563387
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/24/2023
Initial Date FDA Received11/20/2023
Supplement Dates Manufacturer Received11/29/2023
02/21/2024
Supplement Dates FDA Received12/26/2023
03/19/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-