C.R. BARD, INC. (BASD) -3006260740 POWERPORT ISP M.R.I. IMPLANTABLE PORT, CHRONOFLEX SINGLE-LUMEN, 8F; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 1808060 |
Device Problem
Tear, Rip or Hole in Device Packaging (2385)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/15/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: as the lot number for the device was provided, a review of the device history records is currently being performed.The return of the sample is pending.However, a photo was provided for review.The investigation of the reported event is currently underway.H10: d4 (expiration date: 05/2025) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device pending return.
|
|
Event Description
|
It was reported that prior to a port placement procedure, the inner seal of the device was allegedly found to be torn.There was no patient contact.
|
|
Event Description
|
It was reported that prior to a port placement procedure, the inner seal of the device was allegedly found to be torn.There was no patient contact.
|
|
Manufacturer Narrative
|
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the sample was not returned for evaluation.One photo was provided for review.The top label appeared to be partially teared from the tray on the upper left corner.Manufacturing site evaluation of the sample found medical personnel holding the packaging and it was observed in the lower right corner of the tray, that the lid was torn off in this section.No other damage to the packaging is noted.Therefore, the reported tear in device packaging is unconfirmed.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: d4 (expiration date: 05/2025), g3, h6 (method) h11: h6 (result, conclusion) h11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : device not returned.
|
|