• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. LINKED LAG SCREW INSERTER; TRAUMA INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. LINKED LAG SCREW INSERTER; TRAUMA INSTRUMENTS Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/23/2023
Event Type  malfunction  
Event Description
It was reported that the fractured lag screw fractured during a primary surgery.There was no patient harm or impact.Attempts have been made and there is no further information at this time.
 
Manufacturer Narrative
(b)(4).H3: product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
(b)(4).Updated: b4, b5, d2, g3, h1, h2, h3, h4, h6, h10 reported event was confirmed as visual examination of the returned product identified inserter drive shaft is fractured.Wear & tear is seen that indicates repeated use during a potential field age greater than 19 years.Device history record was reviewed and no discrepancies relevant to the reported event were found.A definitive root cause could not be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINKED LAG SCREW INSERTER
Type of Device
TRAUMA INSTRUMENTS
Manufacturer (Section D)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer (Section G)
ZIMMER BIOMET, INC.
1800 w. center st.
warsaw IN 46580
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18168944
MDR Text Key328527757
Report Number0001822565-2023-03193
Device Sequence Number1
Product Code HWR
UDI-Device Identifier00889024036123
UDI-Public(01)00889024036123
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00119301200
Device Lot Number60199460
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/31/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/23/2023
Initial Date FDA Received11/20/2023
Supplement Dates Manufacturer Received02/07/2024
Supplement Dates FDA Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/13/2004
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age16 YR
Patient SexFemale
-
-