Catalog Number LXMC16
Device Problems Migration or Expulsion of Device (1395); Detachment of Device or Device Component (2907); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pyrosis/Heartburn (1883)
Event Date 11/08/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).Date sent: 11/20/2023.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.No lot number was provided therefore a device history could not be done.Photo images were received.Any additional information obtained from the photo evaluation will be included as part of the product investigation.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: what was the date of implant? when did the recurrent gerds begin? what was the date of the imaging which showed the discontinuous linx? what is the product code? what is the device lot number? was the device initially effective in controlling reflux? were any events associated with the onset of symptoms (vomiting, retching, trauma, surgery)? did the patient undergo an mri since device implant? if so, when was the mri and what strength? did the patient have any other surgeries in the area? was any additional imaging performed since device implant? does the device appear to be in a continuous annular state in these images? we are interested in establishing a window when the device may have become discontinuous.Please share any additional images.How was the migration of the device diagnosed (chest x-ray, esophageal line at explant)? are there images available that shows the migration? did the patient have a hiatal hernia at the time of implant? if yes was this repaired at this time of implant? does the patient have a re-occurring hernia now? if yes, did the position of the device change based on the hernia reoccurring? this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that a patient had a linx implanted years ago.She has recurrent gerd and the linx is migrated into the chest.She had an x-ray in february with the device intact.Yesterday she got her pre op ugi and the linx is now disrupted.No change in symptoms.She doesn't quality for another linx because of severe iem.
 
Manufacturer Narrative
(b)(4).Photo analysis: an x-ray image of the device in vivo was reviewed by a medical safety officer.As per medical safety officer: "the images reviewed them straight what appears to be a discontinuous linx device." the mechanism/cause of failure is unknown as a hands-on analysis of the device is necessary to determine the cause of failure.No further investigation can be completed at this point.
 
Manufacturer Narrative
(b)(4) date sent: 4/11/2024 investigation summary a linx device with a visible weld ball that disconnected from a washer was returned to the analysis site.The link length and tensile force measurements were found to meet the applicable specifications during device analysis.The remaining device characteristics, excepting the visible weld ball, show no anomalies for a device that has been reasonably changed as part of the explant procedure, significant tooling marks were noted in some beads.The device was scanned using computer tomography (ct), optical microscopy, and scanning electron microscopy.The washer through-hole at the separation was measured and was greater than the specification.The washer through-hole was concentric with small amount of material displacement at the outer edge of the through hole.The overall appearance of the surface of the washer through hole didn¿t exhibit gross loss of shape.The top view of the diameter of the exposed weld ball was measured.This diameter is within the specification.The weld ball was concentric with the respect to the wire.
 
Manufacturer Narrative
(b)(4).Date sent: 1/3/2024.Additional information received: the product code should be lxmc16.D1, d4.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINX 1.5 16 BEAD US
Type of Device
ANTI-REFLUX IMPLANT
Manufacturer (Section D)
TORAX MEDICAL, INC.
4545 creek road
cincinnati OH
Manufacturer Contact
kate karberg
4545 creek rd.
cincinnati, OH 45242
3035526892
MDR Report Key18169836
MDR Text Key328535203
Report Number3008766073-2023-00232
Device Sequence Number1
Product Code LEI
UDI-Device Identifier00855106005363
UDI-Public00855106005363
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P100049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 04/11/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberLXMC16
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/08/2023
Initial Date FDA Received11/20/2023
Supplement Dates Manufacturer Received12/05/2023
12/20/2023
04/08/2024
Supplement Dates FDA Received12/11/2023
01/03/2024
04/11/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-