The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information and without the device to analyze, a cause of the reported difficult or delayed activation (clip deployment), associated with the inability to confirm separation of clip from the system, could not be determined.The reported retraction problem (excess knob retraction), associated with a more retractable actuator knob, appears to be a normal function of the device.There is no indication of a product quality issue with respect to manufacture, design, or labeling.Na.
|
It was reported that a mitraclip procedure was performed to treat degenerative mitral regurgitation (mr) grade 4.A xtw mitraclip was placed on the lateral commissure of the valve.During deployment, separation of the clip from the delivery system could not be confirmed.Standard troubleshooting was performed and the delivery system alignment was corrected.After about 10 minutes of troubleshooting, the clip separation could be confirmed.The clip was deployed and stable on the leaflets, with mr reduced to grade 1+.The physician noted that the actuator knob was more retractable than usual, and was able to be retracted until the whole 'thick' part was visible.It was not possible to remove the actuator and mandrel so a fracture was ruled out.There was no patient consequences or clinically significant delay.No additional information was provided.
|