• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER GMBH UNKNOWN ACETABULAR COMPONENT; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER GMBH UNKNOWN ACETABULAR COMPONENT; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Difficult to Insert (1316)
Patient Problem Failure of Implant (1924)
Event Date 10/02/2023
Event Type  Injury  
Manufacturer Narrative
(b)(4).D10.Biolox⮠delta, ceramic femoral head, xl, 㸠40/+7, taper 12/14 item#00877504004 lot#3141169 40mm i.D.Size g high wall liner item#30124007 lot#66152419.To date, the device is not returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2023-00639.
 
Event Description
It was reported that a total hip arthroplasty was performed in october of this year and the doctor noticed on x-ray postoperatively that the femoral head was not sitting in the center of the acetabular component.A head and poly exchange was surgically performed and it was found that the polyethylene head was not properly locked into the acetabular component.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.A review of the device manufacturing records for unknown acetabular component could not be performed due to missing lot number.Device is used for treatment.The reported product was reviewed for compatibility with no issues noted.A review of the complaint history could not be performed for unknown acetabular component due to missing reference and lot numbers.Medical records were not provided.With the available information, a definitive root cause could not be determined.No corrective, preventive or field action was taken following the investigation of this event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.Multiple mdr reports were filed for this event, please see associated reports: 0009613350-2023-00639-1.
 
Event Description
Diligence is complete and additional information on the reported event is unavailable at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN ACETABULAR COMPONENT
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ  8404
Manufacturer (Section G)
ZIMMER GMBH
sulzer allee 8
sulzer industrie park
winterthur 8404
SZ   8404
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key18172459
MDR Text Key328557236
Report Number0009613350-2023-00640
Device Sequence Number1
Product Code LZO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
N/A
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/22/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberN/A
Device Catalogue NumberUNKNOWN ACETABULAR COMPONENT.
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/23/2023
Initial Date FDA Received11/20/2023
Supplement Dates Manufacturer Received02/01/2024
Supplement Dates FDA Received02/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient SexMale
Patient Weight137 KG
-
-