A philips product specialist engineer (pse) inspected the device and found visible corrosion on pin 4.Using a fluke hart scientific 1504 thermometer readout the pse performed a temperature test on the batteries which returned results of the batteries was 82 degrees fahrenheit versus ambient temp of 72 degrees fahrenheit.The reporting institution would not disclose further information related to the reported event; however, analysis of the returned device revealed no anomalies related to temperature.Based on this information, the extent of the burn remains unknown at this time, and it does not appear there was any device malfunction which would cause a burn.With the information available and the testing conducted, the cause of the reported problem was not confirmed.The lab tests shows device functioned as intended.The investigation concludes that no further action is required at this time.
|