It was reported that a patient underwent an atrial fibrillation (afib) paroxysmal ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, a force issue was identified.A second device was used to complete the operation.There was no adverse event reported on patient.The force issue is not mdr-reportable; however, the hole in the pebax is.
|
The product investigation was completed.Device evaluation details: the device was returned to biosense webster (bwi) for evaluation.Visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis revealed a hole on pebax and internal parts exposed.The device was connected to carto 3 system, and the device was visualized and recognized correctly; however, error 106 appeared on the system due to an open circuit in the tip area.The hole at the pebax could be related to the issue reported by the customer.The root cause of the hole at the pebax cannot be determined; however, based on the information available, the condition observed most likely was originated in someplace external to the manufacturing environment.A manufacturing record evaluation was performed for the finished device 31071568l number, and no internal actions related to the complaint were found during the review.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: the force sensor of the catheter is disconnected.If the problem persists, replace the catheter cable or the catheter.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|