Catalog Number UNK HIP ACETABULAR CUP PINNACL |
Device Problem
Use of Device Problem (1670)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 11/09/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Revision left total hip replacement for polyethylene wear.Summit size 7 high offset stem, still implanted, 58mm pinnacle cup sector explanted discarded, 58mm 36mm lipped pinnacle liner explanted discarded, 36mm delta ceramic +5 femoral head explanted discarded.It is thought polyethylene wear is likely do to a poorly placed acetabular component.Commented that acetabular component was at a steep inclination increasing stress on polyethylene liner.No further information available.Doi: (b)(6) 2012; doe:(b)(6) 2023; affected side: left.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the device lot number is unknown, therefore a device history review could not be performed. if the lot/serial number becomes available, the record will be re-assessed.
|
|
Search Alerts/Recalls
|
|