H10: the fse replaced the gas delivery system (gds) to resolve the reported issue.The device passed required performance verification tests per philips standards.However, the customer seen the unit and wanted to run the pvt again.During the pvt, the unit failed the leak test.Further troubleshooting was performed, and it was observed that there were additional issues.The boots and white clamps were not installed properly, damage to the pins, boots, o-rings and grommets.The additional issues, troubleshooting and repair will be addressed in separate cases and investigated by the philips complaint handling team.
|