MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION EUROPEAN VALVE SET, EM2400; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938724E
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/13/2023
Event Type  malfunction  
Manufacturer Narrative
E1: initial reporter address: (b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that a part (connector) of a em2400 valve set which connects to a total parenteral nutrition (tpn) bag came out of the rubber tubing.This occurred during bag filling.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added to h3, h4, h6 and h10.H10: the actual device was not available; however, a photograph of the sample was provided for evaluation.The returned photograph was reviewed, and it was observed that the valve set connector was detached from the tubing.Due to the nature of the sample no further testing could be performed.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EUROPEAN VALVE SET, EM2400
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt. 001 mz. 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18178886
MDR Text Key328629826
Report Number1416980-2023-06068
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412475776
UDI-Public(01)00085412475776
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 12/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938724E
Device Lot Number60398457
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/26/2023
Initial Date FDA Received11/20/2023
Supplement Dates Manufacturer Received12/08/2023
Supplement Dates FDA Received12/20/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/22/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
N/A.
-
-