It was reported that on multiple occasions when the patient was charging the vercise genus implantable pulse generator (ipg), the ipg momentarily turned off and on.The patient experienced a shocking sensation during these events.The patient was educated on the charging technique and was provided a new charging system, however the event did not resolve.The database analysis performed identified a bluetooth fault code when charging the ipg.Noise from the charging field of the charger interferes with the bluetooth communication chipset, thus causing internal resets where the device stimulation turns off for 10-15 seconds and then returns to normal operation.No other anomalies were found with the analysis as the device displayed a normal charge log and all impedances were within the expected range.The device remains implanted in the patient delivering stimulation therapy.
|