• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. CYSTO-NEPHRO VIDEOSCOPE Back to Search Results
Model Number CYF-VHA
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device already failed twice the hygiene microbiological investigation (hmi) inspection.The only solution is to replace the possible contaminated parts and perform a third hmi inspection, after repair and final inspection.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
 
Event Description
The customer reported to olympus, the cysto-nephro videoscope tested positive for an unexpected contamination (staphylococcus), and the scoop was damaged at the distal end, and has become a hole.The issues were found during a routine culture of the scope during reprocessing.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Manufacturer Narrative
This supplemental report is to inform that upon further review, this event has already been reported on medwatch 3002808148-2023-08117 (patient identifier # (b)(6) ).Refer to this file for supplemental information related to this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYSTO-NEPHRO VIDEOSCOPE
Type of Device
CYSTO-NEPHRO VIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18181371
MDR Text Key328629990
Report Number3002808148-2023-13155
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeSW
PMA/PMN Number
K062049
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-VHA
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/23/2023
Initial Date FDA Received11/21/2023
Supplement Dates Manufacturer Received01/10/2024
Supplement Dates FDA Received01/18/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-