MPRI ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number AFAPRO28 |
Device Problems
Material Integrity Problem (2978); Protective Measures Problem (3015)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Continuation of d10: product id: 203cx product type: coaxial umbilical cable; product id: 2035u product type: electrical umbilical cable product event summary: the afapro28 balloon catheter with lot number 10297 was returned and analyzed.External visual inspection of the balloon segment showed blood/fluid inside the balloon.The catheter smart chip data was downloaded and reviewed.Data indicated the catheter was used for 18 applications on the reported event date.During functional testing, the console terminated the application and triggered system notice 50006 (the safety system detected blood in the catheter handle, the injection was stopped and the vacuum disabled).Pressurizing of the outer balloon identified the outer balloon proximal bond was leaking and detached from the shaft.In conclusion, the balloon catheter failed the returned product inspection due to a bond detachment at the outer balloon proximal bond.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, a system notice was received indicating that the safety system detected blood in the catheter handle, the injection was stopped and the vacuum disabled.The balloon catheter, electrical umbilical cable, and coaxial umbilical cable were replaced which resolved the issue.The case was completed with cryo.No patient complications have been reported as a result of this event.
|
|
Search Alerts/Recalls
|
|
|