It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, there was ecg (electrocardiogram) signal interference.It was reported that during the operation, the signal interference( noise ) was observed on the intracardiac (ic) channel.During the signal interference, the affected catheter was inside the patient¿s body.The physician still intact ecg (electrocardiogram) signal available to monitor patient heart rhythm.A second device was used to complete the operation.There was no adverse event reported on patient.
|
The product investigation was completed.Device evaluation details: a video was received for evaluation following biosense webster's procedures.The device was returned to biosense webster (bwi) for evaluation. visual inspection and electrical test of the returned device were performed following bwi procedures.Visual analysis revealed reddish material and a hole in the surface of the pebax.An electrical test was performed, and no electrical issues were found.The results were found within the specifications.A manufacturing record evaluation was performed for the finished device 31079977l number, and no internal actions related to the reported complaint condition were identified.Since reddish material was observed inside the pebax, this failure could be related to the noise issue reported by the customer; therefore, the customer complaint was confirmed.The root cause of the pebax damage could be related to the manipulation of the device during the procedure; however, this cannot be conclusively determined.It should be noted that product failure is multifactorial.As part of biosense webster¿s quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: pc-(b)(4).
|