Catalog Number MDE001 |
Device Problems
Material Deformation (2976); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/16/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that there is a damaged thread on the slap hammer.
|
|
Manufacturer Narrative
|
Product complaint #: (b)(4).B3: date of event is an unknown date.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary according to the information received, "damaged thread on slap hammer" the product was not returned to depuy synthes, however photos were provided for review.See attachment (img_1547.Jpg, img_1548.Jpg).The photo investigation revealed that devcie had deformed at the bottom thread.The observed condition was identified as an end of life indicator; damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use, the device must be properly inspected prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions, and inspection procedures.Since the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed as the observed condition of the mde001 would contribute to the complained device issue.Based on the investigation findings, potential cause can be traced to end of life problem, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b3, b5, h6 medical device problem code.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
Additional information received from the sales rep: a.Please clarify what do you mean by damaged? the tip/ thread is worn / damaged.
|
|
Manufacturer Narrative
|
Product complaint (b)(4).Investigation summary: according to the information received, "damaged thread on slap hammer".The product was not returned to depuy synthes.However, photos were provided for review.The photo investigation revealed, that device had deformed at the bottom thread.The observed condition was identified, as an end of life indicator.Damage consistent with repeated use and servicing.The lifecycle requirements of the device are event related and depend on the use and inspection of the device in clinical practice.As the device can be damaged on the first or 100th use.The device must be properly inspected, prior to each surgical use.Refer to the device/country specific ifu for information related to end of life, reprocessing instructions and inspection procedures.Since, the device was not returned, a dimensional inspection cannot be performed.The overall complaint was confirmed.As the observed, condition of the mde001 would contribute to the complained device issue.Based on the investigation findings, potential cause can be traced to end of life problem.And it has been determined, that no corrective and/or preventative action is proposed.There is no indication, that a design or manufacturing issue has caused the complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
|
|