Model Number EGIAUSHORT |
Device Problems
Misfire (2532); Component or Accessory Incompatibility (2897); Firing Problem (4011)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/13/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
D10 concomitant products: egia45amt, egia45amt egia 45 artic med thick sulu (lot#unknown).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
According to the reporter, during a video-assisted thoracoscopic surgical wedge resection procedure, the physician encountered difficulty during the installation of the first reload.The first firing was more strenuous, but ultimately the staples were fully formed.Subsequently, the same handle was used for the installation of the second reload, but the stapling was incomplete.Stapler was then switched with products from another brand.There was no patient injury.
|
|
Manufacturer Narrative
|
Additional information: d9, g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the instrument firing knobs were retracted.The articulation lever was in neutral position.Functional testing found that the instrument loaded, clamped, yet would not cycle due to the sheared firing rack teeth damage.An access hole was cut into the instrument body for visualization of the firing rack.Sheared teeth were visible on the firing rack at site of initial firing post clamp up.It was reported that partial firing and difficult to fire.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.This issue of the sheared teeth on the firing rack may occur in any of the following circumstances: firing over tissue that is beyond the recommended thickness range.Firing with an obstacle incorporated in the jaws.Attempting to fire a reload that is in interlock.In each circumstance, it will become increasingly difficult to actuate the firing handle and the instrument return knobs will be difficult to retract.In addition, staples may not form properly, and tissue may not be fully transected.It was also reported that the device was difficult to load.The reported issue could not be confirmed.The mo st likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: preoperative radiotherapy may result in changes to tissue.These changes may, for example, cause the tissue thickness to exceed the indicated range thickness for the staple size.Careful consideration should be given to any pre-surgical treatment the patient may have undergone and in corresponding selection of staple size.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|