(b)(4).Date sent: 11/22/2023.D4: batch # unk.Additional information was requested and the following was obtained: "there was no change in procedure (e.G.Additional port hole, additional wound) to remove the plastic piece.The patient is stable.The plastic piece will be returning with the device." investigation summary: the product was returned to ethicon for evaluation.Visual inspection was conducted on the returned device.Visual analysis of the returned sample determined that only the sleeve from the b12lt device was received broken.The of plastic from the sleeve was returned inside a plastic bag.In addition, the tyvek was returned along with the instrument.No conclusion could be reached regarding what may have caused the reported incident.One possible cause for the damage found may be due to excessive force being applied to the device.Please refer to the instructions for use for additional information regarding proper device usage.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformances were identified.
|