Model Number 1973-03 |
Device Problem
Entrapment of Device (1212)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/31/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the ranger froze on the guidewire.A ranger paclitaxel-coated pta balloon catheter was selected for use in an angioplasty procedure.A v-18 controlwire was placed without issue.When the physician attempted to advance the ranger delivery system over the v-18, the ranger became stuck and would not advance any further once the balloon began to track over the wire.The ranger device was exchanged in order to complete the procedure.The procedure was completed, and there were no reported consequences to the patient.
|
|
Manufacturer Narrative
|
Device analysis: returned product consisted of a ranger drug coated balloon catheter.The outer shaft, inner shaft, balloon, and tip were visually and microscopically examined.Visual examination revealed buckling to the guidewire lumen 19.5cm from the tip.Microscopic and visual inspection of the remainder of the device presented no other damage or irregularities.Product analysis found damage that would have contributed to the reported event.
|
|
Event Description
|
It was reported that the ranger froze on the guidewire.A ranger paclitaxel-coated pta balloon catheter was selected for use in an angioplasty procedure.A v-18 controlwire was placed without issue.When the physician attempted to advance the ranger delivery system over the v-18, the ranger became stuck and would not advance any further once the balloon began to track over the wire.The ranger device was exchanged in order to complete the procedure.The procedure was completed, and there were no reported consequences to the patient.
|
|
Search Alerts/Recalls
|