Related manufacturer report number: 2017865-2023-73053.It was reported that the patient presented for remote follow-up via merlin.Net.A review of the transmission revealed a decrease in pacing impedance and sensing by the both the right ventricular (rv) and right atrial (ra) leads.Additionally, the rv lead failed to capture.A chest x-ray revealed that both the rv and rv leads were slightly dislocated from their initial positions.Programming interventions were made, and the patient was scheduled for lead revision.During the procedure the physician attempted to reposition the rv, however, the helix failed to retract.The rv lead was explanted and replaced.Further information was requested but not received.
|
The reported events were lead dislodgement, helix mechanism issue, sensing r-wave amplitude variation, and low pacing lead impedance.As received, a complete lead was returned in one piece for analysis.The reported events of sensing r-wave amplitude variation and low lead impedance were not confirmed.Electrical testing did not find any indication of conductor fractures or internal shorts.Visual and x-ray inspection of the lead did not find any anomalies except for procedural damage.The reported event of helix mechanism issue was confirmed.The lead was returned with the helix retracted and clogged with dried blood.X-ray examination found over-torque of the inner coil at the connector region consistent with procedural damage.After the lead was cut and the distal portion was cleaned, torque was then applied directly to the inner coil, and the helix could be extended/retracted.The helix extension length measured within specification.The cause of helix mechanism issue was isolated to dried blood in the helix region, blood on the inner coil, excessive blood at the distal coupling assembly, and over-torqued of the inner coil.
|