The device was returned to olympus for evaluation and the evaluation found the following: the water tightness was lost due to a cut on the bending section cover, the image guide bundle had significant breakages, and the adhesive on the bending section cover had a crack.A review of the device history record found no deviations that could have caused or contributed to the reported issue.The device met all specifications at the time of shipment.It has been over 5 years since the subject device was manufactured.Based on the results of the legal manufacturer's investigation, it is likely that that reported issue occurred due to physical stress on the bending section as it was confirmed that the bending section was broken and the bending section cover was broken; however, a conclusive root cause could not be determined.The user can detect the suggested event by handling device in accordance with the following instructions for use: operation manual_ inspection of the endoscope inspect the external surface of the entire insertion section including the bending section and the distal end for any irregularities such as dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, and protruding objects.The user may be able to prevent the suggested event by handling device in accordance with the following instructions for use: operartion manual_ operation_ precautions_ caution - do not operate the angulation control lever with excessive force in a narrow space to the opposite direction from the bending direction while the distal end of the endoscope is not moved.The bending section may be damaged.Check the tip position of the endoscope and the shape of the bending section using fluoroscopy, etc.Do not insert the insertion tube with excessive force and twist.- do not insert the insertion tube with excessive force into the ureter or calix.The bending section may be damaged.Olympus will continue to monitor field performance for this device.
|
The customer reported to olympus, the fiberscope had black dots in the image.The issue was found during a therapeutic ureteroscopy procedure.The intended procedure was completed using the same set of equipment.The device was returned for evaluation.During the device evaluation, the bending section was broken and was sticking out of the broken metal wires from inside was found.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|