CHRISTOPH MIETHKE GMBH & CO KG PROGAV 2.0 VALVE; HYDROCEPHALUS MANAGEMENT Back to Search Results
Model Number FX410T
Device Problems Mechanical Problem (1384); Infusion or Flow Problem (2964)
Patient Problem Hydrocephalus (3272)
Event Date 11/07/2023
Event Type  Injury  
Event Description
It was reported that a progav 2.0 (#fx410t) was implanted during a procedure performed on (b)(6) 2023.According to the complainant, the valve caused an overdrainage and had adjustment difficulties.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 7 years.Height: 123 cm.Weight: 23 kg.Gender: male.
 
Manufacturer Narrative
Investigation: visual inspection: during the investigation, no significant deformations or damage of the valves was determined.Permeability test: a permeability test has shown that the valve is permeable.Computer control test: to investigate the claim of over- drainage, the opening pressure is measured using a miethke computer control test bench which simulates a cerebrospinal fluid flow.The valve is tested in the horizontal position.The results show that the valve is not operating within the acceptable tolerance in the horizontal position.An accelerated outflow could be confirmed.Adjustment test: the progav 2.0 was tested and is not adjustable.Braking force and brake function test: the brake functionality test has shown that the brake function is operational.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection : after dismantling of the valve, deposits were found in progav 2.0.Result : based on our investigation results, we can determine an accelerated outflow and adjustment difficulties.The determined deposits can be named as the cause for the functional deviations.Organic materila in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 VALVE
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18194686
MDR Text Key328886666
Report Number3004721439-2023-00359
Device Sequence Number1
Product Code JXG
UDI-Device Identifier04041906136529
UDI-Public4041906136529
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K161853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 11/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFX410T
Device Catalogue NumberFX410T
Device Lot Number20045652
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/09/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/09/2023
Initial Date FDA Received11/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/20/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age7 YR
Patient SexMale
Patient Weight23 KG
-
-