MEDTRONIC SINGAPORE OPERATIONS AZURE XT DR MRI SURESCAN; PULSE GENERATOR, PERMANENT, IMPLANTABLE
|
Back to Search Results |
|
Model Number W1DR01 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Pain (1994); Discomfort (2330); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440); Swelling/ Edema (4577)
|
Event Date 10/30/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that approximately twelve days post implantable pulse generator (ipg) implant, and post rotator cuff repair, the patient experienced left arm swelling, redness, pain and tender to palpation over firm cord-like mass area.Thrombophlebitis was suspected so the patient was treated with high does aspirin and celaphexin.The patient then presented to the emergency room with increased swelling and a left upper extremity ultrasound revealed occlusive thrombosis in mid to lateral portion of the left subclavian vein, left axillary vein and left brachial vein with superficial thrombosis in basilica vein over the arm to elbow levels.The patient was then started on rivaroxaban for left, upper extremity deep vein thrombosis.The ipg system remains in use.The patient is a participant in the post approval clinical surveillance product surveillance registry. no further patient complications have been reported as a result of this event.
|
|
Event Description
|
It was further reported that the patient visited the device clinic and visibly reduced swelling was noted with no redness or bruising and the patient denied any pain.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|