It was confirmed that there was a concern by the customer of a delay in recognizing the arrest; however, the concern was not the delay in treatment due to the device, but with the staff response.The complaint was escalated for technical investigation, and the results indicate that the mx450 device log did not show any evidence for a product malfunction on october 23rd, 2023.The audit data from 08:05:27 until 08:26:46 shows a red !!!ecg leads off condition.This inop was silenced at 08:19:48 from the central station, two subsequent reminders were announced at 08:22:48 and 08:26:00.There is also an indication of an asystole alarm at 08:27:02 that was acknowledged/silenced at 08:27:41.Based on the information available and the testing conducted, the cause of the reported problem was user, as the user silenced the alarms.The device was confirmed to be operating per specifications and no failure was identified.If additional information is received the complaint file will be reopened.
|