Reportable based on device analysis completed on 15nov2023.It was reported that a shaft leak occurred.The greater than 70% stenosed target lesion was located in a non-tortuous and non-calcified subclavian artery.An 8.0x40x135 cm express ld vascular stent was advanced for treatment.However, during the procedure, the blood was leaking out via delivery sheath.The device was removed, and the procedure was completed with another of the same device.There were no patient complications reported and the patient status was stable.However, returned device analysis revealed balloon pinhole.
|