Model Number DSX700H11
Device Problems Corroded (1131); Scratched Material (3020)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
H3 other text : device was evaluated by a 3rd party service center.
 
Event Description
The manufacturer received information in relation to a dreamstation auto bipap.There was no report of serious or permanent harm or injury.During evaluation of the device at a third-party service center, bottom enclosure was found to be damaged and ui panel was scratched.The device was found to be functional.Connector oxide contamination was detected.The device was scrapped.At this time, no further investigation can be performed.If any additional information is received, a follow up report will be filed.
 
Manufacturer Narrative
The manufacturer received information in relation to a dreamstation auto bipap unit.The device was returned to a third party service center.During visual inspection of the device, it was determined connector oxide contamination.Additionally, cosmetic defect found as bottom enclosure damage and scratched ui panel during the device evaluation.Device was scrapped at customer's request.Analysis of past events has not indicated an adverse event has occurred due to corrosion.Review of the risk file indicates the potential for a serious adverse event occurring as a result of this incident is unlikely.Additionally, the risk file indicates that corrosion will not substantially affect the performance of the device.This complaint is considered not reportable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DREAMSTATION AUTO BIPAP
Type of Device
VENTILATOR, NON-CONTINUOUS (RESPIRATOR)
Manufacturer (Section D)
RESPIRONICS, INC.
6501 living place
pittsburgh PA 15208
Manufacturer (Section G)
RESPIRONICS, INC.
6501 living place
pittsburgh PA 15208
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18211708
MDR Text Key329039225
Report Number2518422-2023-31785
Device Sequence Number1
Product Code BZD
UDI-Device Identifier00606959022539
UDI-Public00606959022539
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131982
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberDSX700H11
Device Catalogue NumberDSX700H11
Was Device Available for Evaluation? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received11/27/2023
Supplement Dates Manufacturer Received11/07/2023
Supplement Dates FDA Received02/06/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured04/09/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-