Catalog Number 121732058
Device Problem Use of Device Problem (1670)
Patient Problem Insufficient Information (4580)
Event Date 11/01/2023
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
The patient was revised due to dislocation and could not be stabilized during closed reduction.There was no surgical delay.Doi: (b)(6) 2023.Dor: (b)(6) 2023.Affected side: left hip.Additional event information: the cup had more anteversion than usual.The depuy implants had not migrated.Cup was solid.Dual mobility implant was exchanged and patient was stable.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.In order to determine if a lot related issue was possible, a worldwide complaint database search was performed.A worldwide complaint database search found no additional related reports against the provided product code/lot code combination.Based on the inability to find any additional related reports against the provided product code/lot code combination it is reasonable to conclude that there are no anomalies with regard to manufacturing or inspection contained in the device history records that would contribute to the reported event.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PINN SECTOR W/GRIPTION 58MM
Type of Device
PINNACLE HIP SYSTEM : HIP METAL ACETABULAR CUPS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581
3035526892
MDR Report Key18212135
MDR Text Key329064310
Report Number1818910-2023-24036
Device Sequence Number1
Product Code KWA
UDI-Device Identifier10603295010333
UDI-Public10603295010333
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K093646
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 11/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number121732058
Device Lot Number3754347
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/01/2023
Initial Date FDA Received11/27/2023
Supplement Dates Manufacturer Received11/28/2023
Supplement Dates FDA Received11/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/19/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
ALTRX +4 NEUT 36IDX58OD.; DELTA CER HEAD 12/14 36MM +1.5.; PINN CAN BONE SCREW 6.5MMX35MM.
Patient Outcome(s) Required Intervention;
Patient Age57 YR
Patient SexFemale
Patient Weight119 KG
-
-