• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE; ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS WINTER & IBE GMBH HF-RESECTION ELECTRODE; ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL Back to Search Results
Model Number WA22302D
Device Problem Naturally Worn (2988)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/06/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the evaluation found the electrode was broken and not fused, during output, the electrode unexpectedly came into contact with other (metal) instruments, there were foreign objects and impurities attached to the surface of the loop, the liquid surface area in contact with the loop was relatively small, and repeated cutting of the electrodes lead to wear.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the high frequency-resection electrode, loop, was worn out on the distal end.The issue was found during a therapeutic, electroresection of the prostate.The procedure was completed with a similar device.There were no reports of patient harm.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE
Type of Device
ELECTRODE, ELECTROSURGICAL, ACTIVE, UROLOGICAL
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, deutschland 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg 22045
GM   22045
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18216577
MDR Text Key329124594
Report Number9610773-2023-03394
Device Sequence Number1
Product Code FAS
UDI-Device Identifier14042761051665
UDI-Public14042761051665
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K120418
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 11/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA22302D
Device Lot Number1000122980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 11/07/2023
Initial Date FDA Received11/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/03/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-