Catalog Number FNAV-DS-SM |
Device Problems
Difficult to Insert (1316); Difficult to Advance (2920); Material Deformation (2976)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that on 3 november 2023, a small flexnav delivery system was used to deliver a navitor valve.The patient presented with tortuous anatomy.The flexnav was unable to cross the iliac bifurcation on two attempts.The flexnav was removed and the tip was observed to be kinked.A 18f non-abbott introducer sheath was inserted to assist with insertion, but the kinked flexnav could not pass through the introducer.The introducer sheath was changed to a non-abbott 20f and a replacement flexnav with the same navitor valve was able to pass through without issue.There was no patient consequences or clinically significant delay.The patient remained hemodynamically stable throughout the procedure.Patient was reported to be stable.
|
|
Manufacturer Narrative
|
An event of difficulty crossing the iliac bifurcation on two attempts and kink at the tip of flexnav was reported.The delivery system was returned to abbott and the valve capsule was found deformed.The inner membrane was kinked.No other anomalies were noted.Information from field indicated that the patient had tortuous anatomy which may have contributed to the reported event of advancement difficulty and kink at the tip.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.The cause of the reported incident could not be conclusively determined, however the damage noted to the device is consistent with damage during use.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
|
|
Search Alerts/Recalls
|