|
Model Number 1177-01 |
Device Problems
Detachment of Device or Device Component (2907); Difficult to Advance (2920); Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/27/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that device separation occurred.A 5x150mm, 150cm ranger drug coated balloon was selected for a procedure.The balloon was inserted into a 6 french sheath; however, it became stuck in the sheath.Then the physician tried removing the balloon from the patient, the device came apart in two pieces, the catheter and the balloon.Both of the pieces were removed from the patient.There were no patient complications reported.
|
|
Manufacturer Narrative
|
Device evaluation by manufacturer: returned product consisted of a ranger drug coated balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the guidewire lumen is separated 50.1cm from the tip.The inflation lumen is separated 53cm from the tip.There are multiple kinks along the shaft.The inflation lumen and guidewire lumen is stretched.Microscopic examination revealed no additional damages.The hub, proximal end of the inflation lumen and guidewire lumen are missing.
|
|
Event Description
|
It was reported that device separation occurred.A 5x150mm, 150cm ranger drug coated balloon was selected for a procedure.The balloon was inserted into a 6 french sheath; however, it became stuck in the sheath.Then the physician tried removing the balloon from the patient, the device came apart in two pieces, the catheter and the balloon.Both of the pieces were removed from the patient.There were no patient complications reported.
|
|
Search Alerts/Recalls
|
|
|